Dust testing looks like the simplest test in the book: put the sample in a box, add dust, wait. In practice the chamber has three variables that decide the result, and only one of them is the dust.
The dust itself is specified: talcum powder with a defined particle size, used once and then discarded. Reusing dust changes its size distribution and makes the test milder every time.
The three variables
| Variable | Typical setting | What it changes |
| Dust concentration | 2 kg per cubic metre of chamber volume | Too little dust and the test cannot find a path |
| Duration | 8 h for IP5X and IP6X | Long enough for settlement, short enough to be practical |
| Vacuum inside the sample | Up to 2 kPa below ambient, for IP6X only | Turns a passive test into a pressure-driven one |
| Air circulation | Continuous, to keep dust suspended | Without it the dust settles in an hour |
| Dust condition | Dry, sieved, used once | Damp or reused dust gives false passes |
| Sample state | Unpowered unless the standard says otherwise | Powered samples attract dust electrostatically |
IP5X is dust-protected: dust may enter but not in a quantity that interferes. IP6X is dust-tight: no dust at all. The vacuum is what separates them, because it pulls dust through paths that diffusion alone would not find.
Where dust actually enters
- Cable glands without seals. The most common path by a wide margin.
- Screw bosses that break through a wall. A self-tapping screw creates a helical gap.
- Vents and breathers. A membrane rated for water may pass fine dust under vacuum.
- Mating faces without a gasket. Two moulded surfaces touching are not a seal.
- Label edges. A lifted label is a path under vacuum, and a rare one.
Judging the result
For IP6X the judgement is visual: no dust inside, checked immediately after the sample is opened, before anything is wiped. For IP5X the judgement is functional: the sample must still work, and the dust present must not bridge insulation or block a mechanism.
Photograph the inside before touching it. A report that says "small amount of dust" is not a result; a photograph with a scale next to the deposit is. Our dust test chambers include the vacuum pump and the dust-weighing kit, and the technical notes give the chamber volumes that keep the 2 kg/m³ loading practical.
Standards referenced: IEC 60529 clause 13.4 (first digit 5) and 13.6 (first digit 6), IEC 60068-2-68 for the dust and sand methods.
Cleaning between runs is part of the method
Residual dust from the previous run is a different particle size once it has been through the circulation fan. A chamber cleaned every run holds its 2 kg/m³ loading within 5 percent; one cleaned monthly drifts by 20 percent and quietly makes the test easier.
- Weigh the dust in, do not scoop it.
- Discard the dust after every run.
- Check the door seal for dust tracking, which shows as a dark line at 3 mm to 5 mm from the edge.