IPX7 is a defined test: 1 m of water for 30 minutes. IPX8 is not. It is whatever depth and duration the manufacturer declares, agreed with the customer, and greater than the IPX7 condition. That single sentence creates more confusion than any other part of the IP code.
The consequence is that two products both marked IPX8 can be tested to conditions a factor of ten apart, and neither marking is wrong.
From declaration to test condition
| Element | IPX7 | IPX8 |
| Depth | 1 m to the lowest point | Declared by the manufacturer, greater than IPX7 |
| Duration | 30 minutes | Declared, greater than IPX7 |
| Pressure at the seal | About 10 kPa at 1 m | Depth in metres × 9.8 kPa |
| Water temperature | Within a few kelvin of the specimen | Same |
| Acceptance | No harmful ingress | No harmful ingress at the declared condition |
| Marking | IPX7 | IPX8 plus the declared condition in the documentation |
At 1 m the hydrostatic pressure is roughly 10 kPa, which is 0.1 bar, which is a tenth of atmospheric. That sounds small, but it is acting on a seal that was designed to keep air out, not water in, and it is acting for 30 minutes. At 10 m it is 100 kPa, a full atmosphere, and the seal is being pushed in a direction it may never have been analysed for.
The number that actually matters is differential
A sealed enclosure taken from a warm room into cold water sees two pressures at once: hydrostatic and thermal. The air inside contracts as the enclosure cools, creating a partial vacuum that pulls water in through any path. That is why a product can pass a 1 m immersion at equilibrium and fail when it is immersed cold after being stored warm.
Some product standards add a thermal shock element for exactly this reason. If the clause does not, it is still worth recording the specimen temperature at immersion.
Running the test
- Measure depth to the lowest point of the enclosure, not to the water surface above the centre.
- Support the specimen so it is not resting on the tank floor, which shields the bottom surface.
- Start the clock when the specimen is fully submerged, not when it is lowered.
- Dry the outside before opening, or the water on the lid becomes ingress.
- Look for the tide line, which shows where water actually travelled, not just where it pooled.
What to do with a marginal result
A few droplets inside an enclosure with no path to live parts is not automatically a failure, but it is a warning. Record the volume, the position and whether the water reached a creepage path. Then repeat with a longer duration: a seal that lets 0.1 ml in after 30 minutes will let in far more after 4 hours, and the IPX8 declaration is only as good as the duration behind it.
For a product with a cable entry, test with a cable fitted. An immersion tank that takes the enclosure alone will pass a product that fails in service with its gland sealed around a cable. A IP immersion test tank with a depth scale and a specimen support frame keeps the two variables, depth and time, under control, and the site index lists the tanks by diameter and depth.
Standards referenced: IEC 60529, 14.2.8.