IPX5 and IPX6 are both "water jets", which makes them look like two settings on one rig. They are not. IPX5 is a 6.3 mm nozzle at 12.5 litres per minute; IPX6 is a 12.5 mm nozzle at 100 litres per minute, with the pressure held at 100 kPa at the nozzle.
Eight times the flow through a nozzle with four times the area means the supply, the hose and the pump all have to be sized for the larger test. Most rigs that fail to reach IPX6 fail on water, not on pressure.
The two jets side by side
| Parameter | IPX5 | IPX6 |
| Nozzle bore | 6.3 mm | 12.5 mm |
| Flow rate | 12.5 l/min ±5 % | 100 l/min ±5 % |
| Pressure at nozzle | About 30 kPa | About 100 kPa |
| Distance to surface | 2.5 m to 3 m | 2.5 m to 3 m |
| Duration | 1 min per square metre, 3 min minimum | Same rule, longer in practice |
| Water needed per test minute | 12.5 l | 100 l |
| Typical failure of the rig | Pressure drop in a long hose | Pump cannot hold 100 l/min at 100 kPa |
The pressure requirement is measured at the nozzle with the jet running, not at the pump. A 20 m hose at 100 l/min loses a large part of the pump pressure to friction, and the gauge at the pump will happily show 200 kPa while the nozzle sees 90.
What the test is really checking
IPX6 represents heavy seas and powerful washing. The damage it produces is not a slow ingress through a gasket but a direct hit on a seam, a cable gland or a vent. Two design aspects decide the outcome:
- Seam direction. A seam that faces the jet opens under pressure. The same seam facing away stays closed.
- Gland orientation. A cable entry pointing up collects water in the boot; pointing down drains it.
- Vent paths. A pressure-equalising vent that works in IPX5 becomes a water path at three times the dynamic pressure.
- Enclosure flex. A large flat panel deflects under a jet, opening a gap at the fixing points.
Setting up without arguing about the result
Measure the flow with a calibrated meter in line, set the pressure with the jet running, and mark the distance on the floor. Then run the jet over the whole surface at the specified speed, with the sample on a turntable if the standard allows it, so that every part of the enclosure sees the same exposure.
After the test, open the enclosure and look for water where it should not be, not for water in the bottom. A drop on a live terminal is a failure; a puddle under a sealed module is not. Our IP test rigs are supplied with the flow meter and nozzle set sized for both classes, and the technical notes give the hose sizes that keep the pressure drop inside five percent.
Standards referenced: IEC 60529 clause 14.2.5 (IPX5) and 14.2.6 (IPX6), ISO 20653 for the road-vehicle equivalents IPX5K and IPX6K.
Sizing a rig for 100 litres a minute
A 10 min IPX6 test consumes 1,000 litres. A 200 litre reservoir recirculates that water about 5 times per test, which is why the reservoir needs a filter and a temperature check: the pump adds 2 K to 3 K over a 10 min run, and warm water changes nothing electrically but does change the operator experience.
Size the pump for 120 l/min at 120 kPa so the nozzle sees 100 kPa with the hose loss included.